MODERN AND CLEAN DESIGN
RiX70 DC has a super-modern design featuring the flattened generator and transparent cone giving the overall unit the feeling of lightness and clarity.
• Steady and strong support arm.
• Length of the arm 40, 60, 80 and 100 cm.
• Compact and light tubehead.
• Hand switch with 3 mt coiled cord.
• Additional keypad installed near the tube head, for manual adjustment of the exposure time and tooth type.
• Intuitive control panel with quickly identifiable icons.
• Easy and fast installation system.
MORE INFORMATION FOR BETTER RESULTS
The display on the timer shows the exposure parameters, errors and warning messages.
The user-friendly and intuitive design of the icons on the timer allows the user to:
Adjust exposure parameters according to the selected patient size, tooth type and image receptor.
Adapt the radiation dose to lower or higher values.
Modify the exposure time for digital receptor or film.
Select between 60, 65 or 70 kVp.
SYSTEM SUPPLY
Supply Voltage AC100-240V
Frequency 50/60 Hz
Electric current 2.7 A @ 240V , 6.4 A @ 100V
Max Power Consumption 640VA (3 AT fuse)
X-RAY GENERATOR
Tube Voltage 60 kVp ± 8% - 65 kVp ± 8% - 70 kVp ± 8%
Frequency 75 kHz
Anodic current 7 mA± 10%
Output radiation linearity < 0.2 according to IEC 60601-2-63
Duty cycle 1:30
Nominal power 490 W (70 kVp - 7 mA)
Total filtration 1.5 mm Al eq. @ 70 kVp
Focal spot 0.4 mm (IEC 60336)
Leakage radiation at 1 m <0,1 mGy/h
Max thermal capacity of monoblock 130J (convection oil bath cooling)
Beam diameter X Ray at 200 mm 60mm (optional 36 x 45 mm rectangular collimator)
Tube Head Manufacturer Canon Skan X - CEI (Italy)
Model D-045 OX70-4
Inherent filtration < 1 mm Al eq. 0.5 mm Al eq.
Filament current Max. 3 A 1.5 – 2.1 A
Focal spot 0.4 IEC 60336:2005
Target material Tungsten